Preclinical Imaging

M7™ Compact MRI System Specifications

The M7™ compact, high-performance MRI system is designed for dedicated imaging of mice and rats weighing up to 700 grams. The M7™ enables 2D and 3D ex vivo, in vivo and in vitro imaging.

The system’s small footprint and lightweight design enables flexible placement where its needed most — even behind the bio-containment barrier for in vivo imaging.

    Imaging Pulse Sequences

Spin Echo 2D
DWI – with diffusion gradients
IR – with preceding inversion pulse
Fast Spin Echo 2D and 3D
DWI – with diffusion gradients
IR – with preceding inversion pulse
Gradient Echo 2D and 3D
IRSnap GRE based with multiple inversion
time
DCE GRE based Dynamic Contrast
Enhancement
Cardiac Cine GRE based ECG triggered Cine
DTI FSE based diffusion tensor imaging
Dixon FSE based fat separation
Field of View 120 x 120 x 70 mm3
spheroid

 

    Gradients 

Strength 150 (X,Y,Z) mT/m @ 50 A
Linearity
(for Entire
Imaging Volume)
Better than 6.5%
Slew Rate (T/m/s) 750 T/m/s @ 50 A / 200 us

 

 

    Front End — Magnet & Stand

Dimensions
(Height x
Width x Depth)
1320 x 790 x 950 mm /
52 x 31 x 37.5 in
Weight 1550 Kg / 3415 Lbs
Magnet Opening:
Flange Insertion
Diameter
Inner Bore
(Height x Width)

97 mm / 3.8 in

 

220 x 90 mm / 8.7 x 3.5 in

Imaging Volume 120 x 120 x 70 mm3
spheroid
Bº (Tesla) 1.0 T

 

 

    Back End — Electronics Cabinet

Dimensions (Height x
Width x Depth)
1760 x 600 x 1150 mm /
69.3 x 23.6 x 45.3 in
Weight 300 Kg / 600 Lbs

 

   Electrical Requirements

Operator Station Requires an electrical outlet with the
following specifications: 120/230
VAC, single phase, 3 A 50/60 Hz
Line Voltage • 220-240 VAC, 15 A, single phase in
countries using 220-240V
• 208 VAC, 15 A, single phase in
countries using 120V

 

 